null

Insulin Pump Cases

(No reviews yet)
SKU:
AQ158S_0
Condition:
New
Availability:
Ships 1-19 Business Days
$57.06 - $70.97

Delicate insulin pumps and water don’t often mix well. This case will let you bathe, shower or paddle with total peace of mind. 

Key Features

  • Use all buttons and switches through the case
  • Like most of our cases it’ll float if you drop it in the drink
  • Feed out a wire of up to 1/10th inch / 3.5mm diameter from the case
  • The seams are high-frequency welded to form a super-strong bond
  • Waterproofness Warning!: This case features our ‘TC’ Aquaclip which makes the case waterproof to IPX6 (a quick dunking should be fine, but not prolonged immersion).
  • As used by film crews and sound engineers all over the world!

But can I swim with it?


Although it’s not designated as Submersible we have had reports from many, many users that in practice they or their children have enjoyed swimming with our cases with no problem whatsoever (the vast majority of our customers have bought the case for their children). We are talking here about splashing around in the pool or the sea, but we are NOT talking about endurance/long-distance swimming. Our testing shows that the TC clamp will let a tiny amount of water in after a period of total immersion. So the best advice we can give is to wear it in the small of your back if you’re swimming on your front, and on your tummy if you’re swimming on your back, so that it spends as little time submerged as possible. 

Important Notes

  • Cases with this clamp are for surface use only. Very brief immersion (under ten seconds) may be okay but check contents afterwards carefully
  • When not using, leave the clamp open to prevent foam deforming
  • Max allowable thickness of cable is 2.5mm (1/10th inch)
  • Insert belt through loop on the back of case

 Additional Info Regarding Insulin Pumps and Aquapac cases

Aquapac commissioned research by Imperial College London into the suitability of their cases for use with insulin pumps. Results show that in some circumstances insulin flow rate can be reduced by the case. In particular, low flow rates are impeded and as such Aquapac can make no claim regarding the suitability of these cases with insulin pumps. Use of these cases with insulin pumps is entirely at the discretion of the user.

We are advised to recommend that users of Aquapac cases with insulin pumps regularly check their blood sugar levels. Furthermore we strongly recommend that you get the green light from your consultant before use.

The tests were conducted by Imperial College of Science, Technology and Medicine using a Roche Accu-Check Spirit insulin pump at a temperature of 20C at flow rates of 2U/hr and 10U/hr. The amounts of insulin delivered over 60 minutes observed were as follows:

2U/hr not in a case – 0.0131g (flow rate of 100%)
2U/hr in Aquapac 150 – 0.0106g (flow rate of 81%)
2U/hr in Aquapac 150 – 0.0056g (2nd test model)(flow rate of 42%)
2U/hr in Aquapac 550 – 0.0121g (flow rate 92%)

10U/hr not in case – 0.0864g (flow rate of 100%)
10U/hr in Aquapac 150 – 0.0878g (flow rate of 100%)
10U/hr in Aquapac 550 – 0.0903g (flow rate of 100%)